The new range of eVision2 residual gas analyzers with web-enabled technology deliver unrivalled data acquisition speed and quality not possible with previous RGA designs. Patented technology ensures millisecond data point collection even over the full dynamic measurement range and without any loss of data quality. Perfect for rapid leak detection, accurate vacuum diagnostics, chamber diagnostics and contamination monitoring, eVision2 residual gas analyzers provide all the answers...fast.
Integrated, Web-Enabled Residual Gas Analyzers:- Integrated web server, operate your residual gas analyzer and review data from anywhere in the world, or just from your desk.
 - Compatible with all instruments in the MKS process residual gas analyzer range.
- Multiple modes of operation for both process residual gas analysis and general vacuum troubleshooting and diagnostics.
- Operation over any TCP/IP network, including the internet.
Residual Gas Analyzers with Clear Technology Leadership:- Intelligent acquisition that skips larger peaks if requested to do so in order to increase electron multiplier lifetime (download application note here).
- Fastest data acquisition over largest dynamic range of any commercial residual gas analyzer. Performance only previously possible with very expensive ion counting mass spectrometers. Perfect for closed-loop control applications and fast event studies.
|
Advanced RGA Web Browser Applet: | Optional EasyViewTM Application: |  |  |
|
| UHV Surface Analysis Diagnostics: |
|  |
|
Residual Gas Analyzer
| e-Vision2 | | Mass Range | 100 or 200 amu
| | Detector | Faraday only or dual Faraday/SEM
| | Max Operating Pressure | 1E-4 Torr |
| Min Detectable Partial Pressure (3σ baseline noise at 300ms integration) | 2E-11 Torr (Far) 5E-14 Torr (SEM) | | Stability | Better than +/- 0.1 amu over 8 hours | Resolution
| Better than 10% valley throughout mass range | | Bakeout Temperature | 250 degree C | | Mounting Flange | DN-35-CF | Operating Temperature
| 200°C Faraday only with electronics <40°C 150°C Multiplier with electronics <40°C and pressure <1e-8 mbar 90°C Multiplier with electronics <40°C and pressure <1e-5 mbar | | Mass Filter | 100mm |
Residual Gas Analyzer Electronics | e-Vision2 | Mounted Weight on Flange
| 1.7kg | | Power | 24V DC, 2A external universal voltage supply (included) | | Communication | 10/100 Base-T Ethernet, static or automatically assigned IP addresses |
| | Data collection methods | Analogue scanning, full mass range 8, 16 or 32 points/amu Barchart scanning Peak-jump collection of up to 15 peaks per scan | | Maximum data acquisition speed | <3ms per point for analogue scans | Settling time from Full Range to Baseline Signal
| <20ms | | Control method | CE operating system processor with web-server interface | | Filament protection | Opto-isolated input for filament protect or control | Software | Built-in web applications allowing RGA control and data aquisition without installed software using web browser on any PC Optional EasyView for Windows 2000, XP or Vista |
Download Data Sheet: eVision2 Residual Gas Analyzer Data Sheet (492.66 kB) Download Analyser Dimensions: eVision2 Residual Gas Analyzer Dimensional Drawing (699.63 kB)
Related products: custom vacuum systems, custom vacuum chambers, vacuum system controllers, vacuum accessories
We also have a simple Introduction to Mass Spectrometry that might be of interest and there are many more useful resources in our Links section. For additional information about our range of residual gas analysers, if you have any questions, or to request a quotation, please contact us. |